Feature
By the microprocessor all functions are automated, and the
sterilization cycles are monitored. In other words, all phases
such as fill, heat, sterilize, vent, dry and completion are
automatically proceeded. Atany time, the operator may know the
present temperature, the remaining sterilization or drying time,
the proceeding phase and the water level of the reservoir by
means of the display. If there takes place an abnormal
condition, the 6 error messages are given with audible alarm by
the self-diagnosis function. In addition, the door lock system
of the existing screw type has been replaced by the eccentric
latching type. This is a labor-saving, simple and unque device
as the door is sealed if you turn the lever 90.
[ ALARM ]
When an error takes place, 「Err」 is displayed
on the 「Time」 by the self-diagnosis function of
the microprocessor, one of the error numbers (1~6) is displayed
on 「TEMP.」 with the audible signal. [ APPLIANCE
TABLE ] It is the Appliance Table of carbon steel on which the
tabletop sterilizer can be placed with the height and breadth
easiest to use and four casters guaranteeing the excellent
mobility; it is a sectional type easy to store and transport
with elegant surface based on powder coating and strong
durability.
Specification:
|
Overall
Chamber
Reservoir
Tray
Operating Temperature
Air Removal
Power Source
Power Consumption
Control
Time Setting
Press. Gauge
Cycle
Net Weight
|
W413×H340×D548mm
Ø210×L368mm, Cylindrical Type, Capacity :
12.7ℓ
Capacity : 3ℓ
2ea, Stainless St. Plate
121 -132
Gravity Principle
AC230v, 50/60㎐
1,250W
Full Automated by 8 bit Microprocessor
0-60min
Dial Type, Bourdon Tube, 0-4㎏f/㎠
Unwrapped : 132 /5min
Wrapped : 132 /15min, 121 /20min
Liquid : 121 /30min
Dry : 30min
32kg
|
|